TCD Analyzer
A. In Cal Independent screen, user can toggle Cal independent
option between ON and OFF.
I. ON - When the option is ON and the system is single gas
application model, the user must calibrate each range
independently. This is needed when each range is not
linear.
II. OFF - When the option is OFF and the system is in single
gas application model, the user doesn't need to do
independent calibrations. The analyzer will calibrate all
ranges. This is suitable when the application is of a single s
gas and linear on all the ranges that the analyzer is
programmed to have.
4.5.8 Range Polarity
When user clicks on the Range Polarity button in the system screen
2, System shows the Range Polarity screen.
Teledyne Analytical Instruments
Operation
61