Start-up
6.8 System settings
Optimize scanning frequency
The correct signal frequency must now be set ("Horizontal lines. (total)").
1. In order to set "Horizontal lines (total)" correctly, it is best to use an SMPTE and a pixel
on/pixel off test pattern. If the test patterns are not available, you can use an image with
clearly defined border and a line written "lllllll" for the setting. Optimize the signal fre-
quency such that no interferences are present in the picture. If the interval between the
interferences becomes larger, and these therefore also become fewer, you are moving
the slider in the correct direction.
2. If you reach the limit of the control range when setting, you must perform the "Force"
command with the selection "For static pictures" or "For motion pictures". The current
setting is then saved, and the slider set to the center of the control range. Once you
have set the signal frequency optimally, you can continue with the adjustment.
58
SCD 19102
Instructions for Use, 06/2022