Eizo SCD 19102 Instructions For Use Manual page 39

1.3mp 19" lcd monitor
Hide thumbs Also See for SCD 19102:
Table of Contents

Advertisement

Service level 2
Calibration
SCD 19102
Instructions for Use, 06/2022
Function
Adjustment /
setting range
Information
Temperature
Backlight com-
Information
mand
Backlight sensor
Information
info
When backlight
0 ... 100 %
regulation is "On":
Backlight
(regulated)
When backlight
0 - 3360
regulation is "Off":
Backlight
(controlled)
Backlight regula-
Enabled
tion
Disabled
Default: Enabled
ADC calibration
Execute
Backlight margin
Information
6.6 Service level 2 functions
Description
Specified as info in °C
Facility for reading out the current
tab value of the backlight
Backlight sensor regulation
Minimum and maximum regu-
lation
Set brightness of monitor back-
light
Adjustment of overall brightness
to ambient lighting.
Backlight inverter value
The backlight regulation (backlight
sensor) is switched on or off
Automatically calibrate A/D
converter for the applied video
level
The video level range of the sys-
tem is checked, and the monitor
set accordingly. This results in op-
timum adjustment of the individual
RGB A/D converters to the video
source.
The calibration results in a larger
and more flexible video level
range (e.g., the full brightness is
also achieved at 700 mV if the
video level is limited at the value
of 700 mV).
Note: A specific test pattern and
timing are prerequisites!
The A/D converters have already
been factory-set and need not be
readjusted.
Remaining margin of backlight
that can be used for regulation
Note: The data is only displayed
when the panel has warmed up
(after approx. 20 min). The dis-
played value is for information
only.
Start-up
39

Advertisement

Table of Contents
loading

Table of Contents