Connecting
5.7 Serial interface
This serial interface (also called the "service socket") is easily accessible on the back of the
monitor. The 6-pin mini-DIN socket is available for connecting a suitable photometer.
Fig.:
Possible configuration – serial bus mode
If the monitor connection is made via the serial bus, the setting "Serial bus" must first have
been activated using "On Screen Display → Service level 2 → Others". In this case (also
under "On Screen Display → Service level 2 → Others") the bus address has to be as-
signed for each device (1, 2, 3, etc. - assign different numbers for the monitors connected
via the same bus connection).
Settings, measurements, calibration, and automated quality tests can be performed using
QA software.