Mirus for H2/H3 Graingage User's Guide
8.
Tap the Cycle button
Mirus takes its measurements and displays the readings from the sample.
9.
Cycle the sample through the GrainGage (repeat steps 7 and 8) two more times.
This allows you to average the data readings for the sample.
10.
If the calibration file needs to be tuned to match the sample, tap Recalibrate.
Test Weight Recalibration
The test weight recalibration is similar for SCiO, M2.0, and EM.
1.
Open Mirus.
2.
Go to Setup > H3 GrainGage > Moisture Curves or Setup > H2 GrainGage >
Moisture Curves based on your GrainGage model.
3.
Select the calibration you want to modify.
4.
Tap the edit icon
5. Chamber Calibration
.
.
80