This will take you to the moisture calibration screen. For the next steps, see Create a SCiO
Calibration File, Create an M2.0 Calibration File, or Create an EM Calibration File.

Create a SCiO Calibration File

Note: This is only available with the H3 GrainGage and a SCiO sensor installed.
The purpose of the following calibration is to select a crop and calibrate the test chamber
volume (test weight). The default SCiO moisture/protein/oil (where applicable) value in
Mirus is compiled from thousands of samples around the world. While it is possible for
you to offset that data and make it more relevant, HarvestMaster does not recommend
doing so.(If you would like to offset the SCiO moisture/protein/oil (where applicable
see SCiO Offset Recalibration.) 
To begin the calibration you will need one sample that is at least 6 lb (2.7 kg) at the
industry moisture standard.
In the Calibration Wizard,
1.
Enter a Name for the calibration file.
2.
Enter the Chamber inserts count.
The insert count is the number of inserts present in the chamber. It is used to
select default volume for the test chamber, which is important to the accuracy of
the system.
For more information about insert counts and installing the inserts, see the H2
Classic GrainGage
Hardware Manual.
5. Chamber Calibration
64