Mirus for H2/H3 Graingage User's Guide
Temperature is
Out of
Supported
Range
Invalid Scan
Enable or Disable Debug Streaming
This option enables debug streaming messages from the DSP module. This option is on
by default and HarvestMaster recommends leaving it on for troubleshooting purposes.
To find the debug streaming option,
1.
Enter the Mirus Home screen.
2.
Select Setup > H2 or H3 GrainGage > System.
Appendix B: Troubleshooting Mirus
Malfunctioning light source
Sensor temperature exceeds
the allowed range of -10°C to
60°C
Electronic problem resulting
in a spectrum that is not valid
4.
If error persists, run a
self test when
available.
Run a self test when available.
Wait until the sensor
temperature warms (if it was
too cold) or cools (if it was too
hot) and run the sample
again.
1.
Restart the system.
2.
Try to run the sample
again.
3.
If error persists, run a
self test when
available.
157