Manual Calibration Of An Em Moisture Curve Using Excel - Juniper HarvestMaster mirus H2 Classic GrainGage User Manual

Table of Contents

Advertisement

Mirus for H2/H3 Graingage User's Guide
9.
Tap Click to run a sample.
10.
Repeat steps 12–14 with the same sample at least two more times.
Note: The Last Value button can be used to fill the Known Moisture field with the
previous percentage when the same sample is run multiple times.
11.
Tap the next arrow
Mirus displays the adjusted moisture curve with its moisture percentages and
voltages.
Manual Calibration of an EM Moisture Curve Using Excel
HarvestMaster provides a spreadsheet for creating a two-point calibration online. It can
be downloaded here:
2018.xlsx. Follow the instructions in the spreadsheet for creating the curve.
5. Chamber Calibration
to complete the recalibration.
www.harvestmaster.com/data/files/mirus/EM2 Sample Curves
93

Advertisement

Table of Contents
loading

This manual is also suitable for:

Harvestmaster mirus h3 classic graingage

Table of Contents