Mirus for H2/H3 Graingage User's Guide

Test a Calibration File (Moisture Curve)

HarvestMaster recommends testing the new calibration file against different samples to
confirm the data.
1.
Test the grain samples with a reference device by cycling sub-samples from the
larger sample 3–5 times and averaging the data. The averages are the known
values.
2.
Within the hour, take the samples out to the GrainGage. If you are calibrating
outside, make sure the samples stay in the shade.
3.
Open Mirus.
4.
Go to Setup > H2 or H3 GrainGage > Moisture Curves.
5.
Select the moisture curve you want to test.
6.
Tap the edit icon
7.
Pour the grain sample into the hopper (to simulate harvest). Release the isolation
gate so that the grain drops into the GrainGage uniformly.
5. Chamber Calibration
.
79