Mirus for H2 Classic GrainGage
29.
Tap the check icon
If desired, first tap the gear icon
Note: Only adjust these settings if a HarvestMaster Field Service Engineer tells you to.
4.4. 5 Test and Tune a Moisture Cur ve
Note: This section is only applicable to the EM moisture curve.
HarvestMaster recommends testing the new moisture curve against other grain samples to confirm the
moisture and test weight values.
1. Test the grain sample in a benchtop sensor by cycling sub-samples from the larger sample 3–5 times
and averaging the moisture and test weight readings. The averages are the Known Moisture and Test
Weight.
2. Within the hour, take the sample out to the GrainGage. If you are calibrating outside, make sure the
sample stays in the shade.
H2 Classic GrainGage Calibration
to save the new moisture curve.
on the right for more information about this moisture curve.
3. Open Mirus.
4. Go to Setup > H2 GrainGage > Moisture Curves.
5. Select the moisture curve you want to test.
6. Tap the edit icon
.
Gear icon
58