Adjustments / Programming
Overview of ARCOSKOP parameters
•
Call up the "Moreinfo.hlp" help file on the service PC for a short description of the para-
meters available on the ARCOSKOP.
NOTE
Overview of MEMOSKOP Fast parameters
•
The "CDATA.HLP" help file contains a short description of the respective parameters.
The Operating Instructions describe the parameters that can be accessed by the
customer.
Cable module
•
Check the auxiliary voltages at test points TP1= +15 V, TP2= 0V GND,
TP3= -15 V, LED D4 lights up. 24 V at plug ST1.1= +24 V and ST1.2= 0V.
(Tolerance: 22.8 V to 29.5 V).
Holding force after orbital movement of C-arm
•
Voltage at TP6 approx. 9 - 13 V. Adjustable with potentiometer P1.
Sensor
•
Move C-arm to horizontal position, with control unit on top.
•
Switch off orbital brake.
•
Move C-arm approx. 10 cm in clockwise direction.
•
LED D3 lights up at half intensity at TP4 1 V ± 0.1 V. Adjustable by loosening and rotating
the sensor. Following adjustment tighten down lock nut again.
LED D6 motor fault
•
If LED D6 lights up, the motor is then not enabled.
Possible causes: Short circuit or break in sensor, logic control defective.
Siemens AG
Medical Solutions
If parameters have been changed and saved in the "Adjustments,
Parameter" menu with Save to file, you must exit the service pro-
gram with "Logoff" and "Quit" in order to save the modified para-
meters to disk.
SPR2-250.840.01
Rev. 01
Seite 41 von 42
09.05
CS PS 24
5 - 41
5
5
5
5
5
5
ARCOSKOP
System Manual