To do the zero adjustment:
1. Connect the testing leads to the zero-adjust board to create the close to 0 Ω status. See the
section
Appendix 4. Zero
2. Press
3. In the measurement screen, if the measurement range is not in Auto, the following zeroing
operation will zero a single range. Press
execute zeroing, and "Zero Adjusting..." shows at the top right of the screen. As shown in
Figure
39.
4. Once it succeeds, "Zero Adjusted" shows at the top right of the screen. As shown in
If the Instrument zeroed failed, the Instrument briefly prompts Zero adjustment failed, as shown in
Figure
41, and then return to normal measurement.
66
Adjustment.
on the font panel to enter measurement screen.
Figure 39. Zeroing process
on the front panel. The Instrument starts to
Figure 40. Zeroed
Figure
40.