To connect each channel to a cell, connect the SOURCE CH+ and SENSE CH+ to the positive
electrode (+), and connect the SOURCE CH- and SENSE CH- to the negative electrode (-) of the cell.
As shown in
Figure
15.
The channel assignment on each connector of the Multiplexer Card is shown in
Figure 15. Schematic diagram of cell connection - voltage and resistance measurement
The pin definitions of SENSE and SOURCE connectors are the same.
It is recommended to connect to cells using flat twisted pair, shield twisted pair or
coaxial cables.
If a shield twisted pair of shield harness is used, it is recommended to short the
GUARD+ and GUARD- terminals, then connect them to the shield layer.
If a coaxial cable is used, it is recommended to connect the GUARD+ terminal to
shield layers of all signal+'s, and GUARD- to all shield layers of all signal-'s. The
+/- lines of each pair signal of the coaxial should be twisted.
Minimize the length of the non-twisted part.
-
+
Note
Battery Tester
Input and Channel Configuration
Table
10.
39