Fluke BT5300 Series User Manual page 94

Battery tester
Table of Contents

Advertisement

BT5300 Series
User Manual
*OPC
Description
Examples
*OPC?
Description
Examples
*RST
Description
Examples
*SRE
Description
Parameters
Examples
*SRE?
Description
Response
Examples
*STB?
Description
Response
Examples
82
FLUKE,BUND,54010008WS,0
.06,0.04,1.8,0.02,0.02
Set the Bit 0 (Operation Complete) in the Standard Event Status register.
*OPC
Return a "1" to the instrument's output buffer.
*OPC?
Reset the instrument to its power-up configuration, except that the state of IEEE-
488 interface is unchanged.
*RST
Set the Service Request Enable register.
< NR1 (0 to 255)>
*SRE 12
Query the Service Request Enable register.
<NR1 (0 to 255)>
*SRE?
12
Query the Status Byte register.
<NR1 (0 to 255)>
*STB?
12
A "1" is placed in the instrument output buffer.
Binary-weighted decimal value. Each bit
enables or disables the appropriate bit in the
Status Byte register.
Set bit 2 and 3 to one and all other bits to zero.
Bit 2 and 3 are one and all other bits are zero.
Bit 2 and 3 are one and all other bits are zero.