Fluke BT5300 Series User Manual page 52

Battery tester
Table of Contents

Advertisement

BT5300 Series
User Manual
Wiring for Enclosure Potential Measurement
To connect each channel to a cell, connect the SOURCE CH+ and SOURCE CH- to the positive and
negative electrodes of the cell respectively, connect the both corresponding SENSE CH+ and
SOURCE CH- to battery enclosure. As shown in the
Figure 16. Schematic diagram of cell connection - enclosure potential contact check
The enclosure potential contact check can achieve the following tests:
Enclosure potential contact check: detect whether the probe tips connected to the SOURCE
terminals are well connected to the battery enclosure by measuring the resistance.
Positive to enclosure voltage: to measure the voltage between the SENSE CH+ and
SOURCE CH– terminal.
Negative to enclosure voltage: to measure the voltage between the SOURCE CH+ and
SENSE CH– terminal.
The pin definitions of SENSE and SOURCE connectors are the same.
If the Enclosure Potential Contact Check is not needed, and only the the
Positeve/Negative to Enclosure Voltage is measured, the SOURCE CH+ and
SOURCE CH– can be shorted and connected to the battery enclosure.
Table 10.
Pin
Signal
1
CH01+
2
CH01-
3
CH02+
4
CH02-
5
CH03+
6
CH03-
...
...
40
Pinout configuration of the connectors on the Multiplexer Card
Description
Channel 01 +
Channel 01 -
Channel 02 +
Channel 02 -
Channel 03 +
Channel 03 -
...
Figure
16.
-
+
Note
Pin
Signal
35
CH17+
36
CH17-
37
CH18+
38
CH18-
39
CH19+
40
CH19-
41
...
Description
Channel 17 +
Channel 17 -
Channel 18 +
Channel 18 -
Channel 19 +
Channel 19 -
...