Fluke BT5300 Series User Manual page 113

Battery tester
Table of Contents

Advertisement

INITiate:CONTinuous
Description
Parameters
Examples
INITiate:CONTinuous?
Description
Response
Examples
INITiate[:IMMediate]
Description
Examples
TRIGger:SOURce
Description
Parameters
Examples
Enable or disable the continuous measurement. Continuous measurement can
only be set by Remote command. Continuous measurement is enabled when
operation is returned to the Local state or power is ON.
<Boolean>
= OFF or 0
= ON or 1
INIT:CONT OFF
Query the continuous measurement state.
<Boolean>
INIT:CONT?
ON
Transit to the Trigger Wait State. Once the instrument has been configured to
not use continuous measurement (INITiate:CONTinuous OFF), the INITiate
command causes the instrument to take a measurement when the trigger
conditions are met. The last reading can be read by using FETCh? command.
INIT:CONT OFF
INIT
Set the trigger source.
{IMMediate | EXTernal}
= IMMediate
= EXTernal
TRIG:SOUR EXT
Disable continuous measurement. After
measurement, enter the Idle State. Triggering is
ignored in the Idle State. Executing INITiate or
READ? enable the Trigger Wait State.
Enable continuous measurement. After
measurement, enter the Trigger Wait State.
When there is an internal trigger (trigger source
is <IMMediate>), the next trigger is promptly
generated and enters a free run state.
Disable continuous measurement.
Sets the instrument to be triggered by the internal
triggering system.
Sets the instrument to sense triggers through the
trigger key on the front panel of the instrument or
upon on execution of a *TRG command.
Battery Tester
Remote Control
101