 The continuous measurement needs to be enabled to use the storage function.
 Maximum memory storage capacity is 512 measurements. Be aware that attempting to store
more data (by applying a trigger) results in nothing further being stored.
 When the Memory function is enabled, auto-range is not available.
 Memory contents are cleared when performing the following operations:
Enabling the Memory function (OFF to ON)
Changing measurement ranges (ensure all stored measurements in one memory data are in
the same range)
Sending the MEMory:CLEar command
Sending the *RST command
Sending the SYSTem:RESet command
Factory Reset is executed from the UI menu
Power cycling
MEMory:STATe
Description
Parameters
Examples
MEMory:STATe?
Description
Response
Examples
MEMory:CLEar
Description
Examples
Note
Enable or disable the Memory function. Reset to default value (OFF) when
power on.
<Boolean>
= OFF or 0
= ON or 1
MEM:STAT ON
Query the Memory function state.
<Boolean>
MEM:STAT?
ON
Clear the memory data.
MEM:CLE
Battery Tester
Remote Control
97