Fluke BT5300 Series User Manual page 124

Battery tester
Table of Contents

Advertisement

BT5300 Series
User Manual
INITiate
*TRG
STATus:OPERation?
FETCh?
3. Enable continuous initiation, FETCh one measurement
SWITch:MODule INTernal
TRIGger:SOURce EXTernal
ROUTe:CLOSe (@102)
INITiate:CONTinuous ON
*TRG
STATus:OPERation?
FETCh?
Module External, Single Channel, Internal Trigger
In this scenario, it is desired to measure the resistance and/or voltage from the external switch
channel under internal triggering. Below are three methods that can be used.
1. Disable continuous initiation, READ one measurement
SWITch:MODule EXTernal
TRIGger:SOURce IMMediate
ROUTe:CLOSe (@102)
INITiate:CONTinuous OFF
READ?
READ?
2. Disable continuous initiation, FETCh one measurement
SWITch:MODule EXTernal
112
Command
Command
Command
Command
Initiate the instrument to Trigger Wait State.
Trigger one measurement (Trigger Key signal can achieve the
same result).
Query the Operation Statues register until the Bit 11 (Measure
Done) is set.
Return the latest reading (voltage and/or resistance).
Configure the instrument to use the internal switch channel as
measurement input channel.
Configure the trigger source to be external.
Configure channel 102 to be measurement input channel.
Enable continuous measurement.
Trigger one measurement (Trigger Key signal can achieve the
same result).
Query the Operation Statues register until the Bit 11 (Measure
Done) is set.
Return the latest reading (voltage and/or resistance).
Configure the instrument to use the external switch channel as
measurement input channel.
Configure the trigger source to be internal.
Configure channel 102 to be measurement input channel.
Disable continuous measurement.
Acquire one measurement then return the reading (voltage
and/or resistance).
Acquire another measurement then return the reading (voltage
and/or resistance).
Configure the instrument to use the external switch channel as
Action
Action
Action
Action