Item
Module
Selection
Slot
Channel
Function
44
Table 11.
Select whether to use Multiplexer Cards.
- Disable. Not to use any Multiplexer Card, and to measure directly use the
connections on the front panel of the Product.
- Internal. To use the built-in Multiplexer Cards inside the Product.
- External. To use the Multiplexer cards inside the SW1080 Switch Mainframe.
If any Multiplexer Card is selected to be used, this option becomes available, and the
slot where the internal or external Multiplexer Card to be used is in can be selected.
Options include:
- None. No slot is selected.
- 1 to 8. 1 to 2 if the Internal module is selected.
Select the channel number of the corresponding Multiplexer Card, from 01 to 32.
This option is unavailable if no slot is selected.
Select a measurement function
- ACR+DCV
- ACR (only available when Module Selection is Disable)
- DCV (only available when Module Selection is Disable)
- Enclosure Potential Contact Check
- Positive to Enclosure Voltage
- Negative to Enclosure Voltage
For more details to see the section
Channel configuration
Description
Measurement
Function.