Fluke BT5300 Series User Manual page 103

Battery tester
Table of Contents

Advertisement

RESistance:CURRent:MAX
Description
Parameters
Examples
RESistance:CURRent:MAX?
Description
Response
Examples
VOLTage:RANGe
Description
Parameters
Examples
VOLTage:RANGe?
Description
Response
Examples
Set the maximum measurement current. This parameter only applies to the
3 mΩ resistance range.
{ C100 | C200 | C300}
= C100
= C200
= C300
RES:CURR:MAX C300
Query the maximum measurement current.
{C100 | C200 | C300}
RES:CURR:MAX?
C300
Set the voltage measurement range.
<NRf (-10 to 10)>
= [-10, 10]
VOLT:RANG 0
Query the voltage measurement range.
{1.000000E+01 | 1.0000000E+01}
1.000000E+01
1.0000000E+01
VOLT:RANG?
1.000000E+01
The maximum measurement current is 100 mA.
The maximum measurement current is 200 mA.
The maximum measurement current is 300 mA.
The voltage measurement range is 10 V.
Select the voltage measurement range for
measuring 10 V.
The voltage measurement range is 10 V when
the instrument voltage resolution is 6.5-digit.
The voltage measurement range is 10 V when
instrument voltage resolution is 7.5-digit.
Current voltage measurement range is 10 V
when instrument voltage resolution is 6.5-digit.
Battery Tester
Remote Control
91