Fluke BT5300 Series User Manual page 162

Battery tester
Table of Contents

Advertisement

BT5300 Series
User Manual
SAMPle:RATE SLOW
SWITch:MODule:STATe?
INTernal
ROUTe:SCAN (@101:132)
INITiate:CONTinuous OFF
INITiate
STATus:OPERation?
FETCh?
3) Enclosure potential contact check
Determine whether the two probe tips (5, 6) connected to the cell enclosure are conductive
by measuring the resistance, to determine whether they are well connected to the cell
enclosure. If this function test is not required, the connecting wires of probe tips 5 and 6 can
be combined and connected to one probe, for example probe tip 5.
*RST
*CLS
SWITch:MODule INTernal
FUNCtion EPCCheck
RESistance:RANGe 10
SAMPle:RATE EXFAST
CALCulate:AVERage:STATe
OFF
SWITch:MODule:STATe?
INTernal
ROUTe:SCAN (@201:232)
INITiate:CONTinuous OFF
INITiate
150
Command
Command
Set the test speed, such as "Low".
Query the state of the internal Multiplexer Cards. In this
case, return "1, 1" if the states of both SW9010s in slot 1
and slot 2 are normal.
Configure scanning channel: from channel 01 to channel 32
in slot 1.
Disable continuous measurement.
Start scanning.
Query the Operation Status register until the Bit 4 (Sweep
Done) and Bit 8 (Scan Done) is set.
After querying that the corresponding bit in the previous
step is set or waiting until all channel switching and
measurement are completed, send this command to obtain
32 sets of internal resistance (in Ω) and voltage (in V)
readings corresponding to 32 channels.
Reset the Instrument.
Clear registers.
Configure the Instrument to use internal Multiplexer Cards.
Set the instrument function to enclosure potential contact
check.
Set the resistance measurement range to 10 Ω.
Set an appropriate test speed, in this case it is Ex-Fast.
Disable the average function. The purpose of contact detection
is only to determine whether the two probe tips are connected
well, so the averaging function is not needed to improve the
stability of the data.
Query the state of the internal Multiplexer Cards. In this case,
return "1, 1" if the states of both SW9010s in slot 1 and slot 2
are normal.
Configure scanning channel: from channel 01 to channel 32 in
slot 2.
Disable continuous measurement.
Start scanning.
Action
Action