Fluke BT5300 Series User Manual page 130

Battery tester
Table of Contents

Advertisement

BT5300 Series
User Manual
FUNCtion RVOLtage
MEMory:STATe ON
ROUTe:CLOSe (@xxx)
STATus:QUEStionable?
*TRG
STATus:OPERation?
...
MEMory:DATA?
Module Internal / External, Multiple Channel, Internal / External Trigger
*RST
*CLS
TRIGger:SOURce IMMediate
INITiate:CONTinuous ON
SWITch:MODule INTernal
FUNCtion RVOLtage
MEMory:STATe ON
ROUTe:CLOSe (@xxx)
STATus:QUEStionable?
*TRG
STATus:OPERation?
...
MEMory:DATA?
118
Command
Command
as measurement input channel. If the external triggering is
used, the following steps are the same.
Configure to measure the ACR+DCV.
Enable memory function and clear stored memory data.
Configure channel xxx to be measurement input channel.
Query Questionable Status register, the memory function is
triggered successfully only when the Bit 11 (Memory Full) is
not set.
Trigger one storage (Trigger key signal can achieve the
same result).
Query the Operation Status register until the Bit 10 (Memory
Trigger Done) is set.
Repeat the last 4 steps according to measurement
requirement.
Return all the stored memory data.
Reset the instrument to the power-on state.
Clear the Status Byte Register and all event registers, and
clear the Error Queue.
Configure the trigger source to be internal. If the external
triggering is used, the following steps are the same.
Enable continuous measurement.
Configure the instrument to use the internal switch channel
as measurement input channel. If the external triggering is
used, the following steps are the same.
Configure to measure the ACR+DCV.
Enable memory function and clear stored memory data.
Configure channel xxx to be measurement input channel.
Query Questionable Status register, the memory function is
triggered successfully only when the Bit 11 (Memory Full) is
not set.
Trigger one measurement and storage (Trigger Key signal
can achieve the same result).
Query the Operation Status register until the Bit 10 (Memory
Trigger Done) is set.
Repeat the last 4 steps according to measurement
requirement.
Return all the stored memory data.