General Considerations - Agilent Technologies InfinityLab 1290 Infinity II User Manual

Evaporative light scattering detectors
Table of Contents

Advertisement

4
Using the Module

General Considerations

General Considerations

The ELS Detector should be thought of as a detector like any other designed for
liquid chromatography. The main distinguishing feature is the ability to evaporate
the solvent from the column eluent. Therefore, normal system setup precautions
should be remembered when starting to use the instrument. Any solvent that is
intended for use with the ELSD should be fully miscible with any previously used
in the liquid chromatograph. If there is any uncertainty, then a mutually miscible
solvent should be run through the system as an intermediate liquid. Flush the
sample loop also with miscible solvent where necessary. The intended eluent
should be thoroughly degassed, should not contain nonvolatile salts or material,
and should be fully compatible with the columns. All connections should be
made with zero dead volume fittings and tubing with an I.D. ≤0.254 mm
(≤0.010 in).
The ELSD requires nitrogen of purity >98 %, at an inlet pressure of 4.1 – 6.9 bar
(60 – 100 psi). If in-house nitrogen is not available, we recommend the use of a
nitrogen generator with a constant uninterrupted supply of high purity gas. Air
can be used with nonflammable solvent systems. The eluent of choice should be
fully volatile under the chosen detector parameters – any nonvolatilized eluent
will increase baseline noise and reduce sensitivity.
The ELS Detector is a destructive technique and must be placed last when used
in series with other detectors, or used with a flow splitter for semipreparative
applications, where partial sample collection is required.
Agilent InfinityLab ELSD User Manual
55

Advertisement

Table of Contents
loading

Table of Contents