Performance Specifications (G7102A) - Agilent Technologies InfinityLab 1290 Infinity II User Manual

Evaporative light scattering detectors
Table of Contents

Advertisement

2
Site Requirements and Specifications
Performance Specifications

Performance Specifications (G7102A)

Table 6
Type
Light Source
Detector
Nebulizer
Evaporator
Non-cooled
Cooled
Gas Flow Range
Dynamic Range
Short Term Noise
Drift
Operating Pressure
Eluent Flow Range
Digital Output
Remote Operation
Communication
PC Control
Safety and maintenance
In order to control OpenLab CDS edition, the ELSD driver is required.
NOTE
Agilent InfinityLab ELSD User Manual
Performance Specifications G7102A
Specification
LASER 405 nm, 10 mW (Class 3B)
Dual PMT with digital signal processing
OFF, 25 – 90 °C
OFF, 25 – 120 °C
OFF, 10 – 80 °C
0.9 – 3.25 SLM (controlled gas shut-off)
Up to 4 orders of magnitude
<0.1 LSU/h
<1 LSU/h
4.1 – 6.9 bar (60 – 100 psi)
0.2 – 5.0 mL/min
10, 40 or 80 Hz (24 bit)
Remote Start Input
Ethernet
Serial (RS232)
Remote Start Input
Pump Stop: 1 Contact closure
ELSD driver for OpenLab ChemStation
edition
ELSD driver for OpenLab EZChrom edition
LC and CE Drivers Rev. A.02.11 or above
Gas shut-off Valve, Leak Detection, Laser
Interlock
Comment
See conditions in
Table 7
on page 26
See conditions in
Table 7
on page 26