Drying The Diffuser - Agilent Technologies InfinityLab 1290 Infinity II User Manual

Evaporative light scattering detectors
Table of Contents

Advertisement

8
Maintenance

Drying the Diffuser

Drying the Diffuser

If the instrument has been operated incorrectly, the diffuser may become
blocked with liquid. This is manifested by loss of signal, increased baseline noise
and in the extreme case gas bubbling out of the nebulizer drain tube. If this
happens, proceed as follows:
1 Increase the gas flow to 2.8 SLM and increase the evaporator to maximum
2 Reset the instrument to the correct operating conditions and allow to stabilize
Agilent InfinityLab ELSD User Manual
temperature.
The diffuser will be dried out and the instrument ready to use after
approximately 1 h under these conditions.
before continuing.
89

Advertisement

Table of Contents
loading

Table of Contents