7
Error Information
Module Specific Error Messages

Rear Vapor sensor failed

Error ID: 13
The vapor sensor located on the rear panel of the module has failed.
Probable cause
1
2

Vapor detected

Error ID: 14
Probable cause
1
2
3

Leak detected

Error ID: 15
A leak was detected inside the module.
Probable cause
1
2
Agilent InfinityLab ELSD User Manual
Vapor sensor not connected to the
mainboard.
Defective vapor sensor.
External vapor being drawn into unit.
Solvent leak inside unit.
Exhaust tube not fitted.
Loose nebulizer fittings.
Blocked nebulizer causing leak at capillary
fittings.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Remove any source of solvent vapors close to
the module.
Please contact your Agilent service
representative.
Fit black exhaust tube.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
77