Evaporative light scattering detectors
Table of Contents

Advertisement

2
Site Requirements and Specifications
Performance Specifications
Specification Conditions
ASTM: "Evaporative Light Scattering Detectors Used in Liquid Chromatography".
Table 7
Gas flow
Neb temperature
Evaporator temperature
Data Rate
Smoothing
ASTM drift tests require a temperature change below 2 °C/h (3.6 °F/h) over one
hour period. Our published drift specification is based on these conditions. Larger
ambient temperature changes will result in larger drift. Better drift performance
depends on better control of the temperature fluctuations. To realize the highest
performance, minimize the frequency and the amplitude of the temperature
changes to below 1 °C/h (1.8 °F/h). Turbulences around one minute or less can
be ignored. ASTM measurements require that the detector should be turned on
enough time before start of testing.
Agilent InfinityLab ELSD User Manual
Reference conditions
1.6 SLM
70 °C
70 °C
40 Hz
1
26