General Error Messages; Timeout - Agilent Technologies 1260 Infinity User Manual

Fluorescence detector
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

General Error Messages

General error messages are generic to all Agilent series HPLC modules and
may show up on other modules as well.

Timeout

The timeout threshold was exceeded.
Probable cause
1
2
Agilent 1260 FLD User Manual
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Error Information

General Error Messages

Suggested actions
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
Check the logbook for the occurrence and
source of a not-ready condition. Restart the
analysis where required.
7
131