Site Requirements - Agilent Technologies InfinityLab 1290 Infinity II User Manual

Evaporative light scattering detectors
Table of Contents

Advertisement

2
Site Requirements and Specifications

Site Requirements

Site Requirements

A suitable environment is important to ensure optimal performance of the
instrument, see the Site Preparation Checklist for more details.
Laser Safety
The Agilent 1290 Infinity II ELSD (G7102A) is classified as a "Laser Class 1"
product (IEC825-1, CFR1040.10 & 1040.11). During normal operation of the ELSD
no laser light is accessible to the user.
Eye damage by laser light
WAR N IN G
Hazardous laser light can injure eyes.
Do not remove covers and interlocks.
Observe and note the laser warning signs carefully.
Power Considerations
Check the operating voltage of your instrument on the IEC inlet fuse holder on
rear of unit.
Hazard of electrical shock or damage of your instrumentation
WAR N IN G
can result, if the devices are connected to a line voltage higher than
specified.
Connect your instrument to the specified line voltage only.
Inaccessible power plug.
WAR N IN G
In case of emergency it must be possible to disconnect the instrument
from the power line at any time.
Make sure the power connector of the instrument can be easily reached
and unplugged.
Provide sufficient space behind the power socket of the instrument to
unplug the cable.
Agilent InfinityLab ELSD User Manual