Installing the Detector

Site Requirements

A suitable environment is important to ensure optimal performance of the detector.
Power Consideration
The detector power supply has wide ranging capabilities and accepts any line voltage in
the range mentioned in Table 1. Consequently, there is no voltage selector in the rear of
the detector. There are also no externally accessible fuses, because automatic electronic
fuses are implemented in the power supply.
To disconnect the detector from line, unplug the power cord. The power supply still
WARNING
uses some power, even if the power switch on the front panel is turned off.
Shock hazard or damage of your instrumentation can result, if the devices are connected to a line
voltage higher than specified.
WARNING
Make sure to have easy access to the power cable of the instrument, in order to disconnect the
instrument from line.
CAUTION
Power Cords
Different power cords are offered as options with the detector. The female end of all
power cords is identical. It plugs into the power-input socket at the rear of the detector.
The male end of each power cord is different and designed to match the wall socket of a
particular country or region.
14
1100 Series DAD and MWD Reference Manual
Agilent Technologies