AC Voltage Auxiliary Specifications (Dual Output Mode Only)

Range,
Harmonic
Composite
Frequency
Waveform
15 to 45 Hz
45 Hz to 1 kHz
10 to
329.99 mV
1 to 2 kHz
2 to 5 kHz
15 to 45 Hz
45 Hz to 1 kHz
.33 to
3.2999 V
1 to 2 kHz
2 to 5 kHz
15 to 45 Hz
45 Hz to 1 kHz
3.3 to 5 V
1 to 2 kHz
2 to 5 kHz
[1]
For harmonics that are < 1 % of the Fundamental, phase uncertainty is typical.

AC Current Specifications, LCOMP OFF

Range,
Harmonic
Composite
Frequency
Waveform
15 to 45 Hz
45 to 900 Hz
29 to
329.9 μA
900 Hz to 2 kHz
2 to 5 kHz
15 to 45 Hz
45 to 900 Hz
0.33 to
3.299 mA
900 Hz to 2 kHz
2 to 5 kHz
15 to 45 Hz
45 to 900 Hz
3.3 to 32.99 mA
900 Hz to 2 kHz
2 to 5 kHz
15 to 45 Hz
45 to 900 Hz
33 to 329.9 mA
900 Hz to 2 kHz
2 to 5 kHz
[1]
All frequencies can have harmonics up to 100 % of the fundamental; uncertainties are not specified unless otherwise indicated.
[2]
For harmonics that are < 1 % of the Fundamental, phase uncertainty is typical.
Harmonic
Harmonic Amplitude
Amplitude
Uncertainty
Range ( % of
(% of Fundamental + V)
Fundamental)
0.1 to 100 %
0.1 % + 100 μV
0.1 to 100 %
0.1 % + 100 μV
0.1 to 50 %
0.1 % + 100 μV
0.1 to 30 %
0.1 % + 500 μV
0.1 to 100 %
0.1 % + 1 mV
0.1 to 100 %
0.1 % + 1 mV
0.1 to 50 %
0.1 % + 1 mV
0.1 to 30 %
0.1 % + 2 mV
0.1 to 100 %
0.2 % + 3 mV
0.1 to 100 %
0.2 % + 3 mV
0.1 to 50 %
0.2 % + 3 mV
0.1 to 30 %
0.3 % + 3 mV
Harmonic
Harmonic Amplitude
Amplitude
Range
(% of
Fundamental + A)
[1]
Fundamental)
0.1 to 100 %
0.1 % + 0.1 μA
0.1 to 100 %
0.1 % + 0.1 μA
0.1 to 50 %
0.1 % + 0.1 μA
0.1 % + 0.13 μA
0.1 to 30 %
0.1 % + 1 μA
0.1 to 100 %
0.1 % + 1 μA
0.1 to 100 %
0.1 % + 1 μA
0.1 to 50 %
0.1 to 30 %
0.1 % + 1.3 μA
0.1 to 100 %
0.1 % + 10 μA
0.1 to 50 %
0.1 % + 10 μA
0.1 to 30 %
0.1 % + 10 μA
0.1 to 100 %
0.1 % + 13 μA
0.1 to 100 %
0.1 % + 100 μA
0.1 to 100 %
0.1 % + 100 μA
0.1 % + 100 μA
0.1 to 50 %
0.1 % + 130 μA
0.1 to 30 %
5522A PQ Option Specifications
Harmonic Phase
Uncertainty
(Relative to
Fundamental)
0.5 °
1 °
3 °
6 °
0.5 °
0.75 °
2 °
3 °
0.5 °
0.75 °
2 °
3 °
Harmonic Phase
Uncertainty
Uncertainty
(% of
(Relative to
Fundamental)
0.5 °
2 °
3 °
6 °
0.5 °
0.6 °
0.75 °
2 °
0.5 °
0.6 °
0.75 °
2 °
0.5 °
0.75 °
1.5 °
3 °
PQ Option
Absolute RMS
Uncertainty of
Composite Waveform
[1]
(% RMS + V)
0.2 % + 100 μV
0.2 % + 1 mV
0.2 % + 2 mV
Absolute RMS
Uncertainty of
Composite
Waveform
[2]
(% RMS + A)
0.2 % + 0.1 μA
0.2 % + 1 μA
0.2 % + 10 μA
0.2 % + 100 μA
11