Remote Commands And Queries - Fluke 522A/6 Operator's Manual

Multi-product calibrator
Table of Contents

Advertisement

Parameter
Oscilloscope EDGE mode. Programs 25 mV peak-to-peak, 1 MHz, output at the
EDGE
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns EDGE.
Example: SCOPE EDGE; OUT 0.5 V, 5 kHz
Oscilloscope LEVSINE mode. Programs 30 mV peak-to-peak, 50 kHz, output at the
LEVSINE
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns LEVSINE.
Example: SCOPE LEVSINE; OUT 1 V, 50 kHz
Oscilloscope MARKER mode. Programs the period to 1 ms, output at the SCOPE
MARKER
BNC, standby if from OFF or previously in standby. FUNC? returns MARKER.
Example: SCOPE MARKER; OUT 2 MS
Oscilloscope WAVEGEN mode. Programs 20 mV peak-to-peak, square wave, 1 kHz,
no offset, output impedance 1 MΩ, standby if from OFF or previously in standby.
WAVEGEN
FUNC? returns WAVEGEN.
Example: SCOPE WAVEGEN; OUT 1 V, 1 kHz
Oscilloscope VIDEO mode. Programs 100 % output (1 V p-p), line marker 10, format
VIDEO
NTSC. FUNC? returns VIDEO.
Examples:
Oscilloscope PULSE mode. Programs 100 ns pulse width, 1.000 μs period, 2.5 V
PULSE
range. FUNC? returns PULSE.
Example:
Oscilloscope Impedance/Capacitance measurement (MEAS Z) mode. Programs 50 Ω
MEASZ
range. FUNC? returns MEASZ.
Example:
Oscilloscope Overload mode. Programs 5 V dc range. FUNC? returns OVERLD.
OVERLD
Example:
SCOPE
Programs the 5520A-SC1100 oscilloscope calibration option hardware, if installed. The
instrument settings are determined by this command's parameter. Once in SCOPE mode,
use the OUT command to program new output.
Table 10-1. SCOPE Command Parameters (cont.)
Description/Example
(Edge, 0.5 V peak-to-peak, 5 kHz.)
(Leveled sine wave, 1 V peak-to-peak, 50 kHz.)
(Marker, period of 2 ms.)
(Wave Generator, 1 V peak-to-peak, 1 kHz.)
SCOPE VIDEO; OUT 90
(Video, 90 % output)
SCOPE VIDEO; OUT -70
(Video, -70 % output, inverse video)
SCOPE PULSE; OUT 50 ns, 500 ns; RANGE TP8DB
(Pulse, 50 ns pulse width, 500 ns period, 1.5 V range)
SCOPE MEASZ; RANGE TZCAP
(MEAS Z mode, capacitance range)
SCOPE OVERLD; OUT 7 V; RANGE TOLAC
(Overload, 7 V output, ac range)
SC1100 Oscilloscope Calibration Option

Remote Commands and Queries

10
10-29