5080A
Operators Manual
Setting a Dual AC Voltage Output
Tie the terminals NORMAL LO and AUX LO together at the UUT or at the
5080A, via the "LO"s softkey selection "tied."
The calibrator produces a dual ac voltage output by sourcing one ac voltage on the
NORMAL outputs and a second on the AUX terminals.
Complete the following procedure to set a dual ac voltage output. If you make an entry
error, press one or more times to clear the display, then reenter the value.
To avoid damage to the UUT, verify the applied voltage to the
UUT does not exceed the rating of the UUT insulation and the
interconnecting wiring.
1. Press to clear any output from the 5080A.
2. Connect the UUT as described earlier in this chapter under "Connecting the
3. Set the UUT to measure dual ac voltage on the desired range.
4. Press the numeric keys and decimal point key to enter the desired voltage output at
5. Press a multiplier key, if necessary. For example, press .
6. Press .
7. The Control Display now shows the amplitude of your voltage entry. For example,
The AUX output is limited to 5.000 V rms for sine waves.
8. Press the numeric keys and decimal point key to enter the desired voltage output at
9. Press a multiplier key, if necessary. For example, press .
10. Press .
11. The Control Display now shows the amplitude of your entries for the NORMAL
12. Press the numeric keys and decimal point key to enter the desired frequency output
4-24
Calibrator to a UUT."
the NORMAL terminals (maximum five numeric keys). For example, 123.45.
123.45 mV (below).
the AUX terminals (maximum five numeric keys). For example, 234.56.
terminals (upper reading) and AUX terminals (lower reading) (below is typical).
Note
Caution
Note
nn084f.eps
gew367.eps