OPER, STBY, *OPC, *OPC?, and *WAI all operate as described in Chapter 6. The state
of the oscilloscope's output while in SCOPE mode is reflected by the bit in the ISR that
is assigned to SETTLED.
The FUNC? query returns SDCV, SACV, LEVSINE, MARKER, EDGE, and
WAVEGEN for the corresponding oscilloscope modes.
Parameters: OFF
Example:
9-28
Turns the oscilloscope hardware off. Programs 0V,0 Hz,
output at the NORMAL terminals, standby.
Oscilloscope's ac and dc voltage mode. Programs 20 mV
VOLT
peak-to-peak, 1 kHz, output at the
impedance 1 MΩ, standby if from OFF or previously in
standby.
Oscilloscope Edge mode. Programs 25 mV peak-to-peak, 1
EDGE
MHz, output at the SCOPE BNC, standby if from OFF or
previously in standby.
LEVSINE Oscilloscope-leveled sine mode. Programs 30 mV peak-to-
peak, 50 kHz, output at the SCOPE BNC, standby if from OFF
or previously in standby.
Oscilloscope Marker mode. Programs the period to 1 ms,
MARKER
output at the SCOPE BNC, standby if from OFF or previously
in standby.
WAVEGEN Oscilloscope Wavegen mode. Programs 20 mV peak-to-peak,
square wave, 1 kHz, no offset, output impedance 1 MΩ,
standby if from OFF or previously in standby.
SCOPE VOLT;
OUT -2V, 0 Hz
SCOPE VOLT;
OUT 4V, 1 kHz
OUT 0.5V, 5 kHz (Edge, 0.5 V peak to peak,
SCOPE EDGE;
SCOPE LEVSINE; OUT 1V, 20 kHz
OUT 2 MS
SCOPE MARKER;
SCOPE WAVEGEN; OUT 1V, 1 kHz
SCOPE BNC, output
(dc voltage, -2 V)
(ac voltage, 4 V peak-to-
peak, 1 kHz.)
5 kHz.)
(Leveled sine wave, 2 V
peak-to-peak, 20 kHz.)
(Marker, period of 2 ms.)
(Wave Generator, 1 V peak-
to-peak, 1 kHz.)