Ac Voltage (Sinewave) Specifications - Fluke 5500A Service Manual

Multi-product calibrator
Hide thumbs Also See for 5500A:
Table of Contents

Advertisement

5500A
Service Manual
1-8.

AC Voltage (Sinewave) Specifications

Ranges
1.0 to 32.999 mV
33 to 329.999 mV
0.33 to 3.29999 V
3.3 to 32.9999 V
33 to 329.999 V
330 to 1020 V
1-10
Absolute Uncertainty, tcal ± 5 °C
Frequency
90 days
10 to 45 Hz
0.26%
45 Hz to 10 kHz
0.11
10 to 20 kHz
0.15
20 to 50 kHz
0.19
50 to 100 kHz
0.26
100 to 500 kHz
0.75
10 to 45 Hz
0.19
45 Hz to 10 kHz
0.04
10 to 20 kHz
0.08
20 to 50 kHz
0.12
50 to 100 kHz
0.17
100 to 500 kHz
0.53
10 to 45 Hz
0.11
45 Hz to 10 kHz
0.02
10 to 20 kHz
0.06
20 to 50 kHz
0.10
50 to 100 kHz
0.17
100 to 500 kHz
0.38
10 to 45 Hz
0.11
45 Hz to 10 kHz
0.03
10 to 20 kHz
0.06
20 to 50 kHz
0.14
50 to 100 kHz
0.17
45 Hz to 1 kHz
0.04
1 to 10 kHz
0.06
10 to 20 kHz
0.07
45 Hz to 1 kHz
0.04
1 to 5 kHz
0.15
5 to 10 kHz
0.15
± (% of output + µV)
1 year
20 µV
20 µV
0.35%
20
0.15
20
20
0.2
20
20
0.25
20
33
0.35
33
60
1
60
50
0.25
50
20
0.05
20
20
0.1
20
40
0.16
40
170
0.24
170
330
0.7
330
250
0.15
250
60
0.03
60
60
0.08
60
300
0.14
300
1700
0.24
1700
3300
0.5
3300
2500
0.15
2500
600
0.04
600
2600
0.08
2600
5000
0.19
5000
17000
0.24
17000
6.6 mV
0.05
6.6 mV
15
0.08
15
33
0.09
33
80 mV
0.05
80 mV
100
0.20
100
500
0.20
500
Res-
Max
olution
Burden
[1]
1 µV
50 Ω
1 µV
50 Ω
10 µV
10 mA
100 µV
10 mA
1 mV
5 mA,
except
20 mA for
45 to 65 Hz
10 mV
2 mA,
except
6 mA for
45 to 65 Hz

Advertisement