How to Replace the Current Fuses

The current fuses are accessible for the bottom of the Calibrator. These fuses protect the
3A and 20A outputs from over-current. Table 7-2 lists the fuse part numbers for the two
current fuses. To replace a current fuse:
1. Turn the Calibrator over on its top.
2. Remove the two screws with a Phillips head screwdriver as shown in Figure 7-2.
7-4
Table 7-1. Replacement Line Fuses
 109215
 851931
Changing Line Fuse
Changing Line
Figure 7-1. Accessing the Fuse
Fuse Description
5A/250 V Time Delay
2.5A/250 V Time Delay
Voltage
1 2
0
Line Voltage Setting
100 V or 120 V
200 V or 240 V
Line Voltage
Indicator
0 V
( S
B )
1 2
0
gjh004.eps