Returns PT3926 when a 100-ohm RTD with curve α=0.003926 ohms/ohm/°C is set as
the RTD type.
x
IEEE-488
RTD_TYPE_D
(Resistance Temperature Detector Type Default command) Sets the default Resistance
Temperature Detector (RTD) at power on and reset, which is saved in the Calibrator non-
volatile memory. (While saving configuration data in the non-volatile memory, a period
of about 2 seconds, the Calibrator does not respond to remote commands.)
Parameters: PT385
PT385_200
PT385_500
PT385_1000 (1000-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT3926
PT3916
CU10
NI120
Example:
RTD_TYPE_D PT3926
Set the RTD default type to a 100-ohm RTD with curve α=0.003926 ohms/ohm/°C.
RTD_TYPE_D?
x
x
IEEE-488
RS-232
(Resistance Temperature Detector Type Default query) Returns the default Resistance
Temperature Detector (RTD) used at power on and reset.
Responses:
PT385
PT385_200
PT385_500
PT385_1000 (1000-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT3926
PT3916
CU10
NI120
Example:
RTD_TYPE_D? returns PT3926
Returns PT3926 when the RTD default type is a 100-ohm RTD with curve
α=0.003926 ohms/ohm/°C.
x
IEEE-488
SP_SET
(Host Serial Port Set command) Sets the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port, which is saved in the Calibrator non-volatile
memory. (While saving configuration data in the non-volatile memory, a period of about
2 seconds, the Calibrator does not respond to remote commands.) (To set the parameters
for the rear panel SERIAL 2 TO UUT serial port, see the UUT_SET command.) The
factory default values are shown below in bold type. (To return to the factory defaults,
see the FORMAT SETUP command.)
The interface selection sets the command response, with command echo back for
commands and error messages with TERM (terminal) or no echo back with COMP
(computer).
Parameters: <baud rate value>,
x
x
RS-232
Sequential
(100-ohm RTD, curve α=0.00385 ohms/ohm/°C)
(200-ohm RTD, curve α=0.00385 ohms/ohm/°C)
(500-ohm RTD, curve α=0.00385 ohms/ohm/°C)
(100-ohm RTD, curve α=0.003926 ohms/ohm/°C)
(100-ohm RTD, curve α=0.003916 ohms/ohm/°C)
(10-ohm RTD, empirical curve)
(120-ohm RTD, empirical curve)
x
x
Sequential
Overlapped
(100-ohm RTD, curve α=0.00385 ohms/ohm/°C)
(200-ohm RTD, curve α=0.00385 ohms/ohm/°C)
(500-ohm RTD, curve α=0.00385 ohms/ohm/°C)
(100-ohm RTD, curve α=0.003926 ohms/ohm/°C)
(100-ohm RTD, curve α=0.003916 ohms/ohm/°C)
(10-ohm RTD, empirical curve)
(120-ohm RTD, empirical curve)
x
x
RS-232
Sequential
300, 600, 1200, 2400, 4800, 9600
Remote Commands
x
x
Overlapped
x
Coupled
x
x
Overlapped
Coupled
6
Commands
Coupled
6-37

Advertisement

Table of Contents
loading

This manual is also suitable for:

5522a5522a-pq5522a-pq/1g5522a/1ghz5522a-pq/6

Table of Contents