All harmonic specifications below include the fundamental. For waveforms
with no harmonics other than the fundamental, the RMS uncertainty is the
same as the non-PQ mode of the 5520A.
AC Voltage Specifications
Composite
Waveform
Ranges
1 to
32.999 mV
33 to
329.99 mV
0.33 to
3.2999 V
3.3 to
32.999 V
33 to
329.99 V
330 to
1020 V
[1]
All frequencies can have harmonics that are up to 100 % of the fundamental, but uncertainties are not specified unless otherwise
indicated.
[2]
For harmonics that are < 1 % of the Fundamental, phase uncertainty is typical.
[3]
When harmonics of this frequency band are combined with harmonics 45 to 660 Hz, all 45 to 660 Hz harmonics have an uncertainty
of 0.35 % + 25 mV.
[4]
When harmonics of this frequency band are combined with harmonics 45 Hz to 1.2 kHz, all 45 Hz to 1.2 kHz harmonics have an
uncertainty of 0.4 % + 25 mV.
[5]
When harmonics of this frequency band are combined with harmonics 45 to 660 Hz, all 45 to 660 Hz harmonics have an uncertainty
of 0.4 % + 100 mV.
[6]
When harmonics of this frequency band are combined with harmonics 45 Hz to 1.2 kHz, all 45 Hz to 1.2 kHz harmonics have an
uncertainty of 0.5 % + 100 mV.
11-4
Harmonic
Harmonic
Amplitude
Frequency
Range (% of
Fundamental)
15 to 45 Hz
0.1 to 100 %
45 to 900 Hz
0.1 to 100 %
900 Hz to 2 kHz
0.1 to 100 %
2 to 5 kHz
0.1 to 100 %
15 to 45 Hz
0.1 to 100 %
45 to 900 Hz
0.1 to 100 %
900 Hz to 2 kHz
0.1 to 100 %
2 to 5 kHz
0.1 to 100 %
15 to 45 Hz
0.1 to 100 %
45 to 900 Hz
0.1 to 100 %
900 Hz to 2 kHz
0.1 to 100 %
2 to 5 kHz
0.1 to 100 %
15 to 45 Hz
0.1 to 100 %
45 to 900 Hz
0.1 to 100 %
900 Hz to 2 kHz
0.1 to 100 %
2 to 5 kHz
0.1 to 100 %
45 to 440 Hz
0.1 to 100 %
440 to 660 Hz
0.1 to 30 %
660 to 1.2 kHz
0.1 to 10 %
1.2 to 2 kHz
0.1 to 5 %
45 to 440 Hz
0.1 to 100 %
440 to 660 Hz
0.1 to 30 %
660 to 1.2 kHz
0.1 to 10 %
1.2 to 2 kHz
0.1 to 5 %
Note
Harmonic Amplitude
Uncertainty
(% of Fundamental + V)
[1]
0.1 % + 10 μV
0.1 % + 10 μV
0.1 % + 10 μV
0.1 % + 30 μV
0.1 % + 60 μV
0.1 % + 60 μV
0.1 % + 60 μV
0.1 % + 60 μV
0.1 % + 400 μV
0.1 % + 400 μV
0.1 % + 400 μV
0.1 % + 400 μV
0.1 % + 4 mV
0.1 % + 4 mV
0.1 % + 4 mV
0.1 % + 4 mV
0.2 % + 20 mV
0.25 % + 20 mV
[3]
0.35 % + 25 mV
[4]
0.5 % + 40 mV
0.25 % + 100 mV
0.25 % + 100 mV
[5]
0.4 % + 100 mV
[6]
0.6 % + 160 mV
Harmonic
Absolute RMS
Phase
Uncertainty of
Uncertainty
Composite Waveform
(Relative to
(% RMS + V)
[2]
Fundamental)
0.5 °
0.5 °
0.20 % + 6 μV
1 °
3 °
0.5 °
0.5 °
0.20 % + 10 μV
0.8 °
2 °
0.5 °
0.3 °
0.20 % + 100 μV
0.5 °
2 °
0.5 °
0.3 °
0.20 % + 1 mV
0.5 °
2 °
0.75 °
1.2 °
0.20 % + 10 mV
3 °
5 °
0.75 °
1.2 °
0.20 % + 100 mV
3 °
5 °