Verification Tables
The verification test points are provided here as a guide when verification to one-year
specifications is desired.
DC Voltage Verification
Measured Value
Nominal
Value (V dc)
(V dc)
0
0.00125
-0.00125
0.00249
-0.00249
0.0025
-0.0025
0.00625
-0.00625
0.0099
-0.0099
0.01
-0.01
0.0175
-0.0175
0.0249
-0.0249
0.025
-0.025
0.0675
-0.0675
0.1099
-0.1099
0.11
-0.11
0.305
-0.305
Table 10-2. SC1100 Option DC Voltage Verification
(1 MΩ output impedance unless noted)
Deviation
(V dc)
SC1100 Oscilloscope Calibration Option
1-Year Spec. (V dc)
0.00004
0.000040625
0.000040625
0.000041245
0.000041245
0.00004125
0.00004125
0.000043125
0.000043125
0.00004495
0.00004495
0.000045
0.000045
0.00004875
0.00004875
0.00005245
0.00005245
0.0000525
0.0000525
0.00007375
0.00007375
0.00009495
0.00009495
0.000095
0.000095
0.0001925
0.0001925
10
Verification Tables
10-35