Fluke 522A/6 Operator's Manual page 354

Multi-product calibrator
Table of Contents

Advertisement

5522A
Operators Manual
Table 10-7. SC1100 Option Wave Generator Amplitude Verification (1 M output impedance) (cont.)
Wave Shape
triangle
triangle
triangle
triangle
Wave Generator Amplitude Verification: 50
Table 10-8. SC1100 Option Wave Generator Amplitude Verification (50  output impedance)
Wave Shape
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
square
sine
sine
10-40
Nominal
Frequency
Value (V p-p)
(Hz)
0.219
1000
0.899
1000
6.59
1000
55
1000
Nominal
Frequency
Value (V p-p)
(Hz)
0.0018
1000
0.0064
1000
0.0109
1000
0.011
1000
0.028
1000
0.0449
1000
0.045
1000
0.078
1000
0.109
1000
0.11
1000
0.28
1000
0.449
1000
0.45
1000
0.78
1000
1.09
1000
1.1
1000
1.8
1000
2.5
10
2.5
100
2.5
1000
2.5
10000
0.0018
1000
0.0109
1000
Measured
Deviation
Value (V p-p)
(V p-p)
Ω
Output Impedance
Measured
Deviation
Value (V p-p)
(V p-p)
1-Year Spec.
(V p-p)
0.00667
0.02707
0.1978
1.6501
1-Year Spec.
(V p-p)
0.000154
0.000292
0.000427
0.00043
0.00094
0.001447
0.00145
0.00244
0.00337
0.0034
0.0085
0.01357
0.0136
0.0235
0.0328
0.0331
0.0541
0.0751
0.0751
0.0751
0.0751
0.000154
0.000427

Advertisement

Table of Contents
loading

This manual is also suitable for:

5522a5522a-pq5522a-pq/1g5522a/1ghz5522a-pq/6

Table of Contents