Ensure that the calibrator is in standby mode before making
any connection between the calibrator and Tester. Dangerous
voltages may be present on the leads and connectors.
Calibrator Outputs
Normal
Phase
V ac @
in
60 Hz
DEG.
5.0 V
0.0
30.0 mV
8.0 V
0.0
30.0 mV
100.0 V
157.0
150.0 mV
100.0 V
157.0
360.0 mV
10.0 V
46.0
1.40 V
100.0 V
46.0
1.40 V
1. Connect the calibrator to the Model 41 as shown in Figure 4-21.
Voltage is connected to the Model 41 amps channel to simulate current
clamp operation (1 mV = 1 A).
41
POWER HARMONICS
ANALYZER
®
Figure 4-21. Cable Connections for Testing a 40 Series Watts Function
 Warning
Table 4-3. Watts Performance, Text Screen
AUX
W/KW
mV ac
@ 60
MIN
MAX
Hz
145
156
234
246
-14.3k
-13.3k
-37k
-29k
9.2
10.2
92
102
Note
Performance Limits
VA/KVA
VAR/KVAR
Model 41
Only
MIN
MAX
MIN
MAX
145
156
0
4
234
246
0
4
14.5k
15.6k
5.4k
6.3k
32k
40k
10k
18k
13.5
14.5
9.6
10.6
135
145
96
106
Front Panel Operation
Sample Applications
Phase
Harmonics
Screen
MIN
MAX
-2
2
-2
2
155
159
155
159
44
48
44
48
5522A
CALIBRATOR
gjh028.eps
4
4-61