Fluke 522A/6 Operator's Manual page 104

Multi-product calibrator
Table of Contents

Advertisement

5522A
Operators Manual
4-28
WAVE MENUS (Waveform Menus) Opens submenus for selecting the type of
harmonic, waveform, front panel LO terminal condition, and phase.
HARMONIC MENUS (Harmonic Frequency Menus) Opens submenus for
selecting harmonic outputs. See "Setting Harmonics" later in this chapter.
V WAVE (Voltage Waveform) Selects the waveform for the voltage output at the
NORMAL terminals. See "Waveform Types" later in this chapter.
I WAVE (Current Waveform) Selects the waveform for the current output at the
front panel AUX terminals. See "Waveform Types" later in this chapter.
"LO"s (Low Potential Output Terminals) The front panel NORMAL LO and AUX
LO terminals must be tied together either at the UUT or at the Calibrator. When tied
at the UUT, select "open." The default is tied.
Φ & REF MENUS (Phase Difference and 10 MHz reference source.) Selects the
phase difference between the NORMAL and AUX outputs, selects internal or
external 10 MHz reference, and sets the phase difference between an external master
5522A (using 10 MHz (IN/OUT) and the NORMAL output. See "Adjusting the
Phase" and "Synchronizing the Calibrator using 10 MHz ON/OUT" later in this
chapter.
nn088f.eps