Fluke 522A/6 Operator's Manual page 345

Multi-product calibrator
Table of Contents

Advertisement

TRIG
(IEEE-488, RS-232, Overlapped)
Programs the oscilloscope's trigger output BNC.
Parameters: OFF
DIV1
DIV10
DIV100
Example:
TRIG DIV10
TRIG?
(IEEE-488, RS-232, Sequential)
Returns the output setting of the oscilloscope's trigger.
Parameters: (None)
Response:
<character>
OUT_IMP
(IEEE-488, RS-232, Sequential)
Programs the oscilloscope's output impedance.
Parameters: Z50
Z1M
Example:
OUT_IMP Z50
OUT_IMP?
(IEEE-488, RS-232, Sequential)
Returns the impedance setting of the oscilloscope's output.
Parameters: (None)
(Turns the trigger output off.)
(Turns the trigger output on. Frequency is the same as the
signal at SCOPE output.)
(Turns the trigger output on. Frequency is 1/10 of the signal at
SCOPE output.)
(Turns the trigger output on. Frequency is 1/100 of the signal
at SCOPE output.)
(Returns OFF, DIV1, DIV10, or DIV100.)
(Programs the oscilloscope's output impedance to 50 Ω.)
(Programs the oscilloscope's output impedance to 1 MΩ.)
SC1100 Oscilloscope Calibration Option
Remote Commands and Queries
10
10-31