Verification Tests for
AC Current
Range
20.5 A
Wave
IECA
RMS Output 4.8 A
Frequency
50Hz
Table 11-1. PQ Option Verification Table (cont.)
Harmonic
Fundamental
53
20.00%
58
20.00%
63
20.00%
rms
1
100.00%
2
47.00%
3
100.00%
4
18.70%
5
49.60%
6
13.00%
7
33.50%
8
10.00%
9
17.40%
10
8.00%
11
14.30%
12
6.70%
13
9.10%
14
5.70%
15
6.50%
16
5.00%
17
5.80%
18
4.40%
19
5.10%
20
4.00%
21
4.70%
22
3.60%
23
4.30%
24
3.30%
25
3.90%
26
3.10%
27
3.60%
28
2.90%
29
3.40%
30
2.70%
31
3.20%
32
2.50%
33
3.00%
34
2.40%
35
2.80%
36
2.20%
37
2.60%
38
2.10%
39
2.50%
40
2.00%
Amplitude
Specification
Phase
(A)
0
0.3317
0.0133
0
0.3317
0.0133
0
0.3317
0.0133
4.80000
0.0196
2.89500
0.0129
0
1.35900
0.0129
180
2.89500
0.0129
180
0.54123
0.0129
0
1.43500
0.0129
0
0.37760
0.0129
180
0.96918
0.0129
180
0.28950
0.0129
0
0.50347
0.0129
0
0.23160
0.0129
180
0.41536
0.0129
180
0.19300
0.0129
0
0.26432
0.0129
0
0.16543
0.0129
180
0.18880
0.0129
180
0.14475
0.0129
0
0.16659
0.0129
0
0.12867
0.0129
180
0.14905
0.0129
180
0.11580
0.0129
0
0.13486
0.0129
0
0.10527
0.0129
180
0.12313
0.0129
180
0.09650
0.0129
0
0.11328
0.0129
0
0.08908
0.0129
180
0.10489
0.0129
180
0.08271
0.0129
0
0.09766
0.0129
0
0.07720
0.0129
180
0.09136
0.0129
180
0.07237
0.0129
0
0.08582
0.0129
0
0.06812
0.0129
180
0.08092
0.0129
180
0.06433
0.0129
0
0.07654
0.0129
0
0.06095
0.0129
180
0.07262
0.0129
180
0.05790
0.0129
PQ Option
Verification Table
Specification
(A)
(deg)
3.0
3.0
3.0
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
0.6
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0
1.0