Figure 5-6. Testing the RS-232 UUT Port via IEEE-488 Port
Complete the following procedure to test RS-232 UUT port operation via the IEEE-488
port using the Win32 Interactive Control utility.
1. Complete the "IEEE-488 Port Setup Procedure" earlier in this chapter to set up the
Calibrator for GPIB operation.
2. Complete "Testing the IEEE-488 Port" to prepare the Calibrator IEEE-488 port for
testing. Before the final step, return to this procedure and continue to Step 3 below.
3. Go to Start then to the Programs menu.
4. Select "NI-488.2M software for... (your operating system)".
5. From the NI488.2M software menu, select "Win32 interactive control".
6. A DOS window opens with a prompt as shown here:
7. At the prompt, type the following line to activate the IEEE interface card:
<ibdev 0 4 0 10 1 0>
8. The second number in this line is the primary address of the calibrator. If the address
has been changed from the factory default, change this line accordingly.
How to Set up the RS-232 UUT Port for Remote Control
Modem Cable
IEEE-488 Cable
Controller
Remote Operations
SERIAL 2
TO UUT
Port
5522A Calibrator
gjh046.eps
5
5-19