Table 7-5. Verification Tests for DC Current (AUX) (cont.)
Range
2.99999 A
2.99999 A
2.99999 A
2.99999 A
20.5000 A
20.5000 A
20.5000 A
20.5000 A
20.5000 A
Table 7-6. Verification Tests for Resistance
Range
10.9999 Ω
10.9999 Ω
10.9999 Ω
32.9999 Ω
32.9999 Ω
32.9999 Ω
109.9999 Ω
109.9999 Ω
329.9999 Ω
329.9999 Ω
329.9999 Ω
1.099999 kΩ
1.099999 kΩ
3.299999 kΩ
3.299999 kΩ
3.299999 kΩ
10.99999 kΩ
10.99999 kΩ
32.99999 kΩ
32.99999 kΩ
32.99999 kΩ
Output
1.09000 A
-1.09000 A
2.99000 A
-2.99000 A
0.0000 A
10.9000 A
-10.9000 A
20.0000 A
-20.0000 A
Output
0.0000 Ω
2.0000 Ω
10.9000 Ω
11.9000 Ω
19.0000 Ω
30.0000 Ω
33.0000 Ω
109.0000 Ω
119.0000 Ω
190.0000 Ω
300.0000 Ω
0.330000 kΩ
1.090000 kΩ
1.190000 kΩ
1.900000 kΩ
3.000000 kΩ
3.30000 kΩ
10.90000 kΩ
11.90000 kΩ
19.00000 kΩ
30.00000 kΩ
Lower Limit
1.08979 A
-1.09021 A
2.98906 A
-2.99094 A
-0.0005 A
10.8954 A
-10.9046 A
19.9833 A
-20.0168 A
Lower Limit
-0.0010 Ω
1.9989 Ω
10.8986 Ω
11.8982 Ω
18.9980 Ω
29.9978 Ω
32.9979 Ω
108.9962 Ω
118.9954 Ω
189.9938 Ω
299.9914 Ω
0.329991 kΩ
1.089974 kΩ
1.189954 kΩ
1.899938 kΩ
2.999914 kΩ
3.29991 kΩ
10.89974 kΩ
11.89954 kΩ
18.99938 kΩ
29.99914 kΩ
Maintenance
Performance Tests
Upper Limit
1.09021 A
-1.08979 A
2.99094 A
-2.98906 A
0.0005 A
10.9046 A
-10.8954 A
20.0168 A
-19.9833 A
Upper Limit
0.0010 Ω
2.0011 Ω
10.9014 Ω
11.9018 Ω
19.0020 Ω
30.0023 Ω
33.0021 Ω
109.0038 Ω
119.0046 Ω
190.0062 Ω
300.0086 Ω
0.330009 kΩ
1.090026 kΩ
1.190046 kΩ
1.900062 kΩ
3.000086 kΩ
3.30009 kΩ
10.90026 kΩ
11.90046 kΩ
19.00062 kΩ
30.00086 kΩ