Fluke 522A/6 Operator's Manual page 250

Multi-product calibrator
Table of Contents

Advertisement

5522A
Operators Manual
Range
32.9999 V
32.9999 V
32.9999 V
329.999 V
329.999 V
329.999 V
329.999 V
329.999 V
329.999 V
329.999 V
329.999 V
1020.00 V
1020.00 V
1020.00 V
1020.00 V
1020.00 V
1020.00 V
1020.00 V
1020.00 V
[1]
Typical specification is -24 dB at 2 MHz
Range
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
329.999 mV
7-12
Table 7-7. Verification Tests for AC Voltage (Normal) (cont.)
Output
Frequency
30.0000 V
30.0000 V
30.0000 V
33.000 V
33.000 V
300.000 V
300.000 V
300.000 V
300.000 V
300.000 V
200.000 V
330.00 V
330.00 V
1000.00 V
1000.00 V
1000.00 V
1000.00 V
1020.00 V
1020.00 V
Table 7-8. Verification Tests for AC Voltage (AUX)
[1]
Output, AUX
Frequency
10.000 mV
10.000 mV
10.000 mV
10.000 mV
10.000 mV
300.000 mV
300.000 mV
300.000 mV
300.000 mV
300.000 mV
300.000 mV
Lower Limit
20 kHz
29.9928 V
50 kHz
29.9904 V
90 kHz
29.9759 V
45 Hz
32.993 V
10 kHz
32.989 V
45 Hz
299.953 V
1 kHz
299.953 V
10 kHz
299.946 V
18 kHz
299.928 V
50 kHz
299.922 V
100 kHz
199.630 V
45 Hz
329.91 V
10 kHz
329.91 V
45 Hz
999.74 V
1 kHz
999.79 V
5 kHz
999.79 V
8 kHz
999.74 V
1 kHz
1019.79 V
8 kHz
1019.74 V
Lower Limit
45 Hz
9.622 mV
1 kHz
9.622 mV
5 kHz
9.535 mV
10 kHz
9.520 mV
30 kHz
8.700 mV
9.5 Hz
283.350 mV
10 Hz
299.180 mV
45 Hz
299.390 mV
1 kHz
299.390 mV
5 kHz
299.100 mV
10 kHz
298.650 mV
Upper Limit
30.0072 V
30.0096 V
30.0241 V
33.007 V
33.011 V
300.047 V
300.047 V
300.054 V
300.072 V
300.078 V
200.370 V
330.09 V
330.09 V
1000.26 V
1000.21 V
1000.21 V
1000.26 V
1020.21 V
1020.27 V
Upper Limit
10.378 mV
10.378 mV
10.465 mV
10.480 mV
11.300 mV
316.650 mV
300.820 mV
300.610 mV
300.610 mV
300.900 mV
301.350 mV