Emd Information - A&D TM-2441 Instruction Manual

Recorder for ambulatory blood pressure monitor
Hide thumbs Also See for TM-2441:
Table of Contents

Advertisement

11.2. EMD Information

The requirements that apply to medical electronic instruments are
described below:
Performance within the EMD guidelines
The use of the recorder require special precautions for EMD
(Electromagnetic Disturbances). Operate the recorder in accordance
with the warnings for EMD described in this manual. Portable and
mobile RF communication equipment (e.g. cell phones) can affect
medical electrical equipment.
Accessories compliant with EMD standards
The accessories and options for this recorder meet the conditions of
IEC60601-1-2:2014. If an unauthorized accessory is used, it may
cause increasing emissions and lower noise immunity.
Use accessories designated by the A&D company.
Unauthorized accessories may be influenced by
electromagnetic emission and have reduced immunity
against disturbances.
EMISSIONS LIMITS
Radiated RF emission
Phenomenon
CISPR11
Warning
96
Compliance
Group 1, Class B

Advertisement

Table of Contents
loading

Table of Contents