A&D TM-2441 Instruction Manual page 73

Recorder for ambulatory blood pressure monitor
Hide thumbs Also See for TM-2441:
Table of Contents

Advertisement

#1
: When the sleep mode is set to "ON", the start time and operation
time of the automated measurement and the interval time of the
sleep mode can be used. The interval time for these sections
(1 to 6) cannot be used.
: When the sleep mode is set to "OFF", the interval time for the sleep
#2
mode is not displayed.
: If the start time is specified and the operation time is set to
#3
"OFF", when the AUTO switch is set "ON", the automated
measurement starts at the preset start time and continues
until the AUTO switch is set to "OFF". If the AUTO switch
is set to "ON" again, the automated measurement starts at
the preset start time.
When the operation time is specified, even if the AUTO
switch is operated during the automated measurement, the
automated measurement continues for the operation time
from the time that the AUTO switch is initially set to "ON".
#4
: If the start time is set to "OFF" and the operation time is
specified, when the AUTO switch is set to "ON", the
automated measurement performs the first blood pressure
measurement and continues for the operation time.
If the AUTO switch is set to "OFF" during the automated
measurement, it stops.
If the AUTO switch is set to "ON" again, the automated
measurement is performed for the operation time.