9.2. Cleaning the Product

Clean the recorder before use and after use.
Clean the recorder before attaching to the next patient.
Do not spray water and immerse in water to clean the recorder.
This may cause a malfunction.
Dry the recorder after wiping it with water and the antiseptic
solution, so that the liquid does not penetrate the recorder.
Disinfect the recorder periodically to ensure infection prevention.
Do not use a sterilizer on the recorder.
Do not use an organic solvent (for example: thinner) or
povidone-iodine solution to clean the recorder. It may cause
discoloration, damage and malfunction.
Do not use a hair dryer etc. to dry the recorder. It may cause
malfunction and damage.
Confirmation after Cleaning the Cuff
Confirm that the cuff bladder is correctly inserted inside the cuff
cloth. If it is not correctly inserted, damage or explosion may
occur during inflation.
Cleaning the recorder
Wipe the dirt and dust from the exterior case of the recorder using
gauze moistened with water or warm water and squeezed well.
When blood or medicines, etc. adhere to the case, firstly, clean with
gauze moistened with antiseptic solution and squeezed well. Then,
wipe the wet case using gauze moistened with water or warm water
and squeezed well. We recommend using chemicals (ingredient
name) from the antiseptic solution listed in the table (Example of
useable antiseptic solution (Ingredient name)).